Tel: 86-0731-83889629
EN
sve kategorije

Početna> Proizvodi > Općenito i instrument za analizu ulja

KONTAKTIRAJTE NAS

Kanta za led

Tel: 86-0731-83889629

Mob: 86 13548954520

E-mail: [e-pošta zaštićena]

Skype: changshafriend.tester

WhatsApp: 86 13548954520

astm_d93_fdt_0292_automatic_closed_cup_flash_point_tester
ASTM D93 FDT-0292, Automatic Closed Cup flash point tester

ASTM D93 FDT-0292, Automatic Closed Cup flash point tester

Automatic closed cup flash point tester is conformed to GB/T261, ASTM D93 and the latest ISO 2719:Determination of flash point -- pensky-martens closed cup method.

UPIT
  • Specifikacija
  • Konkurentska prednost

Mjesto podrijetla:

Changsha, Kina

Brand Name:

PRIJATELJ

Broj modela:

FDT-0292

Certifikacija:

CE, ISO, CO

Minimalna narudžba Količina:

1 SET

Cijena:

8000–15000 USD

Pakiranje Detalji:

Tvrdi karton, standardno izvozno pakiranje

Vrijeme isporuke: (VODEĆI)

30 DANA

Uvjeti plaćanja:

T / T, L / C

Nabava Sposobnost:

50SET/MJESEC

Primjena:
1. ASTM D93
2. Ispitivanje benzina, ulja za podmazivanje, loživog ulja
3. Nafta, energija, kemija, istraživanje, kontrola kvalitete i sveučilište

Tehnički podaci:

Tehnički parametri

Napajanje

AC 220V ± 10%, 50Hz

Raspon

Sobna temperatura ~ 300 ℃

Rezolucija

0.1 ℃

 

Agitation speed

90( r./min)~120(r./min).    (GB/T 261 procedure A,ASTM D93  procedure A)

250(r/min)±10(r/min).   (GB/T261 procedure B,ASTM D93  procedure B)

Brzina grijanja

5(℃ / min ) ~6(℃ / min )   ( GB/T 261 procedure A,ASTM D93 procedure A)

1(℃ / min )~1.5(℃ / min)( GB/T 261 procedure B,ASTM D93 procedure B )

 

Paljenje

Electronic ignition and the gas flame

To adapt to the ambient temperature

10-40 ° C

To adapt to the environment humidity

Ukupna potrošnja energije

Manje od 500W

svojstvo

1. The instrument uses A&B testing process for different types of sample selection. It can also custom set the agitation speed

2. The instrument uses the annular integral heating structure for more uniform heating and more accurate heating rate

3. The Instrument uses the air-cooling temperature control system to ensure the furnace body temperature is constant.

4. The instrument use a self-diagnostic function, instrument failure automatically prompt function

5. The instrument uses a self-over-temperature protection, over-temperature range can be set

6. The instrument uses the LCD touch display, real-time simulation tracking display temperature and test time function curve, the operation interface prompt the real time function

7. The instrument uses the voltage limit technology in Friend’s proprietary intellectual property rights, to ensure the constant of platinum wire heating and improve the instrument testing accuracy

8. The instrument uses the precision machining technology, which is conformed to the structural standard in ASTM D93

9. The instrument uses two kinds of heating devices: electronic ignition and the gas flame, which is convenient and faster.

123jpg3-14567

Kontaktirajte Nas